 +====== AN012 - How to reset battery alarm on C1-455-B. ======
 +
 +  *Get the programming cable built as shown:
 +{{:strumenti:qmove:q1:an013:cavo_prog_rs232.png?nolink&600|}}
 +  *Replace the battery with the **instrument switched on**.
 +  *Install the program [[https://www.qemservice.it/download/Prog/Other/ttermp23.zip|"Teraterm Pro"]] after the extraction and follow the installation instructions.
 +  *Open the "Teraterm Pro".
 +  *When opening the following screen will appear. Set the desired COM port.
 +
 +{{:strumenti:qmove:c1455:an012:1.png?nolink&400}}
 +
 +  *Clock on the "Setup" menu and then select "Serial port". Setting the speed to **19200**.
 +  *Put in boot the C1-455 instrument, by pressing the "RESET" system button and now the "BOOT" button.
 +  *If the procedure is correct led "BOOT" lights.
 +  *Open the "Control" menu and select "Macro". From the explorer window, select the file [[http://www.qemservice.it/download/Prog/Other/ver.zip|"ver.ttl"]] after extraction.
 +
 +{{:strumenti:qmove:c1455:an012:2.png?nolink&400}}
 +
 +  *At this point the following screen will appear. Select "Yes".
 +
 +{{:strumenti:qmove:c1455:an012:3.png?nolink&400}}
 +
 +  *Set the date and follow the directions.
 +  *At this point the "BATT" led turns off.