AN013 - Battery allarm reset procedure for Q1-CPU-B (modular Qmove).
Obtain cable built as shown:
The battery should be substitute only with
power on instrument
.
To install the program
"Teraterm Pro"
after extraction and follow the installation instruction.
Open the “Teraterm Pro”.
When you open the following screen appears. To set the COM port.
Click on “Setup” menù and than select the “Serial port”. To set the speed at
19200
.
Put the CPU in boot state, push “RESET” system button and immediately the “BOOT” button.
If the procedure is correct “BOOT” led turn ON.
Open the “Control” menù and select “Macro”. From the window of the explorer, to select the file
"ver.ttl"
after extraction.
The following screen appears. To select “Yes”.
Set the “data” and then follow the instructions.
The “BATT” led turn OFF.