Differences
This shows you the differences between two versions of the page.
Next revision | Previous revision | ||
en:appnote:an018 [2017/05/05 17:31] – created qem103 | en:appnote:an018 [2019/10/30 09:15] (current) – old revision restored (2017/05/05 17:43) qem202 | ||
---|---|---|---|
Line 1: | Line 1: | ||
+ | |||
+ | ====== Reset checksum error procedure on R201A ====== | ||
+ | |||
+ | The purpose of the following document is to integrate the procedure already described in the Hardware Manual of the [[en: | ||
+ | |||
+ | ==== R201 instrument normally in operation ==== | ||
+ | {{: | ||
+ | |||
+ | ==== QC043 instrument usually in operation ==== | ||
+ | {{: | ||
+ | |||
+ | |||
+ | < | ||
+ | ==== FUNC and BOOT keys ==== | ||
+ | {{: | ||
+ | |||
+ | ====System LEDs==== | ||
+ | {{: | ||
+ | |||
+ | |||
+ | < | ||
+ | ==== Procedure for error checksum reset of the r201==== | ||
+ | |||
+ | * Instrument in error CHECKSUM: led **PWR** fixed, led **ERR** makes two flashes. | ||
+ | {{: | ||
+ | |||
+ | * Restart the instrument by pressing and holding the **FUNC** key | ||
+ | {{: | ||
+ | |||
+ | * So you enter in the system functions: led **PWR** on and led **L4** on. | ||
+ | {{: | ||
+ | |||
+ | * Press the **BOOT** key for 2 seconds. | ||
+ | {{: | ||
+ | |||
+ | * Start the error reset procedure: led **PWR** flashes. | ||
+ | |||
+ | |||
+ | * End the error reset procedure and the instrument restarts. | ||
+ | |||
+ | |||
+ | * NOTE: If, to end the **L4** and **ERR** LED lights up fixed manually restart the instrument. | ||
+ | {{: | ||
+ | |||
+ | * To the end the R201 instrument return in run state | ||
+ | {{: | ||
+ | |||
+ | |||
+ | < | ||
+ | ==== Error message checksum from QC043 ==== | ||
+ | |||
+ | * If the R201 is in checksum error then in the QC043 will appear the following screen: | ||
+ | {{: | ||
+ | |||
+ | * Press the **F2** key on the panel to continue. | ||
+ | {{: | ||
+ | |||