AN034 - Application update procedure with MMC for J1-P20
Document: | AN030 | ||
---|---|---|---|
Description: | Application update procedure with MMC for J1-P20 | ||
Editor: | Riccardo Furlato | ||
Approver | Gabriele Bazzi | ||
Link: | http://www.qem.eu/doku/doku.php/appnote/an034 | ||
Language: | English | ||
Document Release | Description | Note | Date |
01 | New manual | 27/01/2020 |
Premise
Before switching to the application update from MMC there are some premises to be respected:
-
The MMC must be formatted with a filesystem type FAT12, FAT16 or FAT32;
-
Before copying the files to the MMC, it is recommended that you format it.
-
The file must be named as:
appqtp.bin
Procedure
-
Turn off the instrument.
-
Press and hold the
F1
button to access the SYSTEM FUNCTIONS. -
Turn on the instrument.
-
A successful operation pops up the message SUCCESSFUL to indicate that the operation is successful.
-
Press the
F1
button to restart the system.